semi合集-English.pdf - 第6296页

SEMI G59-94 © SEMI ,1994, 2002 3 11.1.3 Store the multi-ing r edient standard solution an d the calibration solutions in the correct flask s and label with the ion type and concentration. NOT E 5: New Sta ndard solutions…

100%1 / 7923
SEMI G59-94 © SEMI 1994, 2002 2
Table 1 Sensitivity of Ion Chromatograph
Ion Sensitivity(ng/mL)
Na
+
0.2
NH
4
+
0.5
Cation
K
+
1.0
Cl
-
0.3
PO
4
3-
2.0
Br
-
1.0
NO
3
-
1.0
Anion
SO
4
2-
1.0
8.2 Chart Recorder
8.3 Ion Extraction Vessels — Polypropylene or
polytetrafluoroethylene containers with sealing caps.
NOTE 1: The contamination level of these vessels must be
less than one fifth (1/5) of the expected contamination level
on the interleaf of leadframes when measured in a control test.
8.4 Water Bath — 300 mm L × 300 mmW × 200
mmH, filled with DI water, and capable of holding
95°C.
8.5 Constant Temperature and Humidity Chamber
8.6 Volumetric Dispenser — (e.g., Pipettes — 10 mL
and 100 mL capacity).
8.7 Quartz Flasks and Pipettes for Cation Standard
Solutions100, 250, 500, and 1000 mL capacity
(flasks); 1, 10, and 25 mL capacity (pipettes).
8.8 Borosilicate Glass Flasks and Pipettes for Anion
Standard Solutions — 100, 250, 500, and 1000 mL
capacity (flasks); 1, 10, and 25 mL capacity (pipettes).
8.9 Chemical Balance, Weighing Chemicals
8.10 Scissors, Tweezers, Spatula
9 Reagents and Materials
9.1 Deionized water, resistivity 15 megohm
centimeters at 25°C per ASTM D 1193.
9.2 Eluents and regenerants for specific chromatograph
columns prepared per chromatograph equipment
manufacturer’s recommendations so that the water peak
can be separated from the ionic peaks.
9.3 Compounds Required for the Preparation of
Standard Solutions
9.3.1 Cations — NaCl, NH
4
Cl, KCl.
9.3.2 Anions — NaCl, Na
2
HPO
4
•12H
2
O, NaBr,
NaNO
3
, K
2
SO
4
.
NOTE 2: All compounds must be reagent grade.
10 Sampling
10.1 Sample Conditioning
10.1.1 In case of measurement of ionic contamination
transferred from the interleaves to the leadframes, select
a stack of ten (10) leadframe strips with their nine (9)
interleaves alternately, from the lot to be tested, and
place them horizontally into a chamber at 85 ± 5°C, 85
± 5% Relative Humidity for 24 hours.
10.2 Sample Selection
10.2.1 Recommended sample size of interleaf used in
interleaf extraction test is 10,000 sq. mm.
10.2.2 In the extraction procedure, do not use the top or
bottom leadframe strips in the stack.
NOTE 3: If a load is used to hold the stack together, it shall
be recorded as part of the conditions of test.
NOTE 4: The leadframe/interleaf contact area shall be
recorded. The vender and customer shall agree on the surface
area of the leadframes.
11 Preparation of Standard Solutions
11.1 Standard Solutions
11.1.1 The single ingredient standard solutions of each
ion (Na
+
, NH
4
+
, K
+
, Cl
-
, NO
3
-
, Br
-
, SO
4
2-
, PO
4
3-
)are
made by dissolving 1.000 g of each ion into 1.000 liter
of DI water, respectively. The stored multi-ingredient
standard solutions shown in Table 2 are then made from
these single ingredient standard solutions by the
dilution method.
Table 2 Concentration of Standard Solution for
Calibration
Standard Solution for
Calibration(ng/mL)
Ion Mixed Standard
Solution
g/mL)
I II III
Na
+
10 5 10 20
NH
4
+
10 5 10 20
Cation
K
+
10 5 10 20
Cl
-
4 10 20 40
PO
4
3-
10 25 50 100
Br
-
4 10 20 40
NO
3
-
4 10 20 40
Anion
SO
4
2-
4 10 20 40
11.1.2 Cation and Anion standard solutions for
calibration are made by diluting the stored multi-
ingredient standard solutions as shown in Table 2.
SEMI G59-94 © SEMI ,1994, 2002 3
11.1.3 Store the multi-ingredient standard solution and
the calibration solutions in the correct flasks and label
with the ion type and concentration.
NOTE 5: New Standard solutions for calibration are required
every 24 hours. Ensure that the flasks are cleaned with water
before refilling with a new solution.
11.2 Calibration
11.2.1 Set up the chromatograph and regenerate the
columns according to the manufacturer’s instructions
(ASTM D 4327 provides further details).
11.2.2 Run the eluent through the chromatograph until
a stable baseline chromatograph is obtained.
11.2.3 Select the injection volume recommended by
the manufacturer for each ion type and inject it into the
chromatograph. Record the chromatograph for each ion
type, and make the calibration curve for each ion (ion
concentration versus peak height or area).
NOTE 6: Peak height or area under the ion’s characteristic
curve is proportional to the concentration.
12 Procedure — Container Extraction Method
NOTE 7: The size of the extraction vessels depends on the
expected volume of water and the leadframes. The vessels
must be at least threequarters filled with water and
leadframes.
NOTE 8: The vessels must be from the same manufacturing
batch.
12.1 Extraction Vessels — Cleaning
12.1.1 Fill three vessels three-quarters of the way full
in order to reduce the amount of air in the vessel and
attach the caps.
12.1.2 Place the vessels in a water bath at > 95°C for
30 ± 2 minutes.
12.1.3 Remove the vessels from the bath and rinse out
five (5) times with DI water.
12.2 Interleaf Contamination Extraction
12.2.1 Place the interleaf sample in one of the cleaned
vessels.
NOTE 9: The interleafing may be cut in order to ease loading
into the vessels.
12.2.2 Add 100 mL of DI water and cap the vessels.
Place a similar amount of water into the other cleaned
vessel and attach the cap.
NOTE 10: 100 mL is the recommended volume of water;
however, the samples must be covered with water.
12.3 Leadframe Contamination Extraction
12.3.1 Place five (5) leadframe strips in one of the
vessels cleaned per Section 12.1.
NOTE 11: The leadframes may be cut as required to ease
entry into the vessel.
12.3.2 Cover with DI water and cap.
NOTE 12: 25 mL is the recommended volume of water for a
small volume of leadframe samples, 100 mL for a large
volume. The samples must be covered with water.
12.3.3 Place a similar volume of water in one of the
vessels cleaned per Section 12.1.
12.4 Extration
12.4.1 Place the three (3) vessels into the water bath at
T 95°C, for 30 ± 2 minutes.
12.4.2 Remove the vessels from the bath and allow to
cool.
12.4.3 Remove the leadframes and interleafing
material from their respective vessel and recap.
13 Measurements and Calculations
13.1 Chromatograph Preparation and Calibration
13.1.1 Prepare the chromatograph for operation by
regenerating the columns according to the
manufacturer’s recommendations.
13.1.2 Run the eluent through the chromatograph until
a stable baseline calibration is established.
13.2 Testing
13.2.1 Inject the recommended sample size of solution
from the interleaf extraction into the chromatograph
and obtain the chromatogram.
13.2.2 Repeat 13.1 and then inject the recommended
sample size from the leadframe extraction and obtain
the chromatogram.
13.2.3 Repeat 13.1 and then inject the recommended
sample size from the water-only vessel and obtain the
chromatogram of the background sample.
NOTE 13: The time from extraction to insertion of the sample
into the chromatograph shall not exceed eight (8) hours.
13.3 Results
13.3.1 Sample concentrations are determined from the
calibration curves for each ion type.
13.3.2 The surface concentration of ionic contaminants
(SCIC) for each ion type is given by the following
equation:
SEMI G59-94 © SEMI 1994, 2002 4
SCIC ng/cm
2
()
=
Sample
Concentration
Background
Concentration
æ
è
ç
ö
ø
÷ ×
Extraction
Volume(mL)
Total Interleaf(or Leadframe) Surface Area(cm
2
)
14 Report
The report, when used by a vender to certify a user’s
requirement, or by a user at incoming inspection, shall,
at least, contain the following information. Additional
information shall be agreed between user and supplier.
14.1 Vendor’s lot numbers for leadframes and interleaf
material, and date of shipment.
14.2 Sample conditioning conditions.
14.3 Test conditions.
15 Related Documents
15.1 SEMI Specifications
SEMI G52 — Standard Test Method for Measurement
of Ionic Contamination on Semiconductor Leadframes
(Proposed)
NOTICE: These standards do not purport to address
safety issues, if any, associated with their use. It is the
responsibility of the user of these standards to establish
appropriate safety and health practices and determine
the applicability of regulatory limitations prior to use.
SEMI makes no warranties or representations as to the
suitability of the standards set forth herein for any
particular application. The determination of the
suitability of the standard is solely the responsibility of
the user. Users are cautioned to refer to manufacturer’s
instructions, product labels, product data sheets, and
other relevant literature respecting any materials
mentioned herein. These standards are subject to
change without notice.
The user’s attention is called to the possibility that
compliance with this standard may require use of
copyrighted material or of an invention covered by
patent rights. By publication of this standard, SEMI
takes no position respecting the validity of any patent
rights or copyrights asserted in connection with any
item mentioned in this standard. Users of this standard
are expressly advised that determination of any such
patent rights or copyrights, and the risk of infringement
of such rights, are entirely their responsibility.
Copyright by SEMI ® (Semiconductor Equipment and Materials
International), 3081 Zanker Road, San Jose, CA 95134. Reproduction o
f
the contents in whole or in part is forbidden, without express writte
n
consent of SEMI.