SEMI P36-0600 © SEMI 2000 1 SEMI P36-0600 GUIDELINE OF MAGNIFICATION REFERENCE FOR CRITIC AL DIME NSION MEASUREMENT SC A NNI NG ELECTRON MICROSCOPES (CD-SEMS ) This g uideline was technically approved by the Global Me tr…

100%1 / 7923
SEMI P35-0704 © SEMI 2000, 2004 8
can be made consistent with the ultimate purpose of the
measurements.
A1-4 The inner and outer linewidth bounding boxes
need not have the same shape, but their widths must be
unambiguously defined. A rectangular shape for both
will result in a centerline, centroid, and linewidth
independent of height above the substrate, but may also
result in a larger interval between the inner and outer
linewidth bounding boxes. The figure 95% represents 2
standard deviations of the Gaussian probability
distribution of possible values for the linewidth. The
multiplier value of 2 (the “coverage factor”) is also
often used even if the distribution is not Gaussian
[reference Section 3.2], but then the interpretation of
95% confidence no longer applies. Three standard
deviations or some other multiple may be used if so
specified.
A1-5 The probability distribution and expectation
value for the position of an edge within the line edge
bounding box are determined as described in ANSI
Z540-2 [reference Section 3.2]. Default values for these
assume the edge is equally likely to be anywhere inside
the line edge bounding box. In that case the expectation
value of the line edge location is the center of the line
edge bounding box (i.e., the linewidth is the width of
the mean linewidth bounding box [Section 5.8.3]), and
the edge position uncertainty (at the 95% confidence
level) is 0.577 × width of line edge bounding box. The
corresponding linewidth measurement uncertainty
(Section 5.8.7) component is 0.816 × width of line edge
bounding box if the right and left edge location
uncertainties are uncorrelated, and 1.154 × width of line
edge bounding box if they are mirror-image correlated
(as is often approximately the case). See reference
Section 3.2, page 13.
A1-6 In most cases, the width or centroid or edge
positions of the bounding box is measured from its
image in a metrology tool; inferring the width of the
bounding box from this image usually requires
modeling of the image-forming process. The bounding
box should be constructed so that its image in the
metrology tool can be modeled with the modeling tools
available. If the image is not modeled accurately,
additional measurement uncertainty will accrue
[reference Section 6.4].
NOTICE: SEMI makes no warranties or
representations as to the suitability of the standards set
forth herein for any particular application. The
determination of the suitability of the standard is solely
the responsibility of the user. Users are cautioned to
refer to manufacturer' s instructions, product labels,
product data sheets, and other relevant literature,
respecting any materials or equipment mentioned
herein. These standards are subject to change without
notice.
By publication of this standard, Semiconductor
Equipment and Materials International (SEMI) takes no
position respecting the validity of any patent rights or
copyrights asserted in connection with any items
mentioned in this standard. Users of this standard are
expressly advised that determination of any such patent
rights or copyrights, and the risk of infringement of
such rights are entirely their own responsibility.
Copyright by SEMI® (Semiconductor Equipment and Materials
International), 3081 Zanker Road, San Jose, CA 95134. Reproduction o
f
the contents in whole or in part is forbidden without express written
consent of SEMI.
SEMI P36-0600 © SEMI 20001
SEMI P36-0600
GUIDELINE OF MAGNIFICATION REFERENCE FOR CRITICAL
DIMENSION MEASUREMENT SCANNING ELECTRON MICROSCOPES
(CD-SEMS)
This guideline was technically approved by the Global Metrology Committee and is the direct responsibility
of the Japanese Metrology Committee. Current edition approved by the Japanese Regional Standards
Committee on March 1, 2000. Initially available at www.semi.org May 2000; to be published June 2000.
1 Purpose
1.1 The purpose of this guideline is (1) to define
common and important specifications of magnification
references which are used for calibrating magnifications
of critical dimension measurement scanning electron
microscopes (CD-SEMs), and as the result (2) to
provide magnification references which are easy for
anyone to use.
2 Scope
2.1 It is preferable that design, manufacture and
purchase specifications for CD-SEM magnification
references conform to this guideline.
2.2 This standard does not purport to address safety
issues, if any, associated with its use. It is the
responsibility of the users of this standard to establish
appropriate safety and health practices and determine
the applicability of regulatory limitations prior to use.
3 Referenced Standards
3.1 None.
4 Terminology
4.1 CD-SEM magnification refere nce
a CD-SEM
magnification reference is defined as a standard for
calibrating magnifications of a said CD-SEM through
mounting the standard on the specimen stage,
measuring the dimensions of reference patterns formed
on the standard, determining the difference between the
measurement value and the true or reference value of
the reference patterns, and adjusting the CD-SEM
parameters to bring the difference zero.
4.1.1 There are two types of CD-SEM magnification
references: one is a wafer on which reference patterns
are formed; another is a chip on which reference
patterns are formed. Hereinafter, the former will be
referred to as “wafer-type magnification reference,” and
the latter will be referred to as “chip-type magnification
reference.”
4.2 edge roughness Edge rough ness refers to edge
variations seen in the SEM images, and is defined as
the distance, within a field of view, between the peak
line and the valley line, where the peak line means the
line which runs through the highest peak and is parallel
to the pattern-edge mean line, the valley line means the
line which runs through the lowest valley and is parallel
to the pattern-edge mean line, and the pattern-edge
mean line conforms to the expected pattern-edge line.
Here, a peak is the tip of a convex section, and a valley
is the deepest part of a concave section (See Figure 1).
5 Ordering Information
5.1 Specify whether a said CD-SEM magnification
reference is a wafer-type magnification reference or a
chip-type magnification reference.
6 Requirements
6.1 CD-SEM Magnification References and Their
Structure
6.1.1 CD-SEM Magnification References are either
wafer type or chip type. Chip types are composed of a
chip (or chips) and a stub to hold the chip (or chips).
Wafer type include a 'drop-in' wafer which is a wafer
with a thinned section where a die with the reference
calibration pattern can be dropped in.
6.2 Shape, Size and Mounting Method of CD-SEM
Magnification References
6.2.1 Shape and size of wafer-type magnification
references, and method of mounting them on the
specimen stages in CD-SEMs must follow the shape,
size and mounting method of product wafers.
6.2.2 Shape and size of chip-type magnification
references, and method of mounting them on the
specimen stages of CD-SEMs are defined as follows:
6.2.2.1 Size and shape of chip-type magnification
references: See Figure 2.
6.2.2.2 Mounting method of chip-type magnification
references:
6.2.2.2.1 A magnification reference should be
mounted on a specimen stage as markers made on the
stub fit in markers made on the specimen stage, and be
fixed by using a screw. The markers made on the stub
should show the direction of reference patterns at an
accuracy of within ± 1°.
SEMI P36-0600 © SEMI 2000 2
6.2.2.2.2 The markers made on the specimen stage
should show the X- and Y-direction of specimen-stage
movement, and the angle between the two markers
should be 90° ± 0.1° (See Figures 2 and 3).
6.2.2.3 Chip surface height for chip-type
magnification references:
6.2.2.3.1 The height of chip surface (or reference
pattern surface) from the specimen stage surface should
be 1.7 ± 0.05 mm as the magnification reference is
mounted on the specimen stage.
6.3 Materials for Magnification R eferences
Materials for magnification references must be non-
magnetic and conductive.
6.3.1 Wafers for wafer-type magnifi cation references
and chips for chip-type magnification references are
preferable to be silicon.
6.3.2 Materials of the stub for chip-t ype magnification
references are preferable to be aluminum, copper, or
carbon.
6.4 Properties of Magnification Reference Pattern
6.4.1 Reference patterns do not change or degrade in
use.
6.4.2 The block area of reference patterns (or every
block area, in the case where there are some reference
pattern blocks) must be larger than 100 µm × 100 µm,
and the block should be placed so that it is easily
recognizable.
6.4.3 The pattern-edge roughness must be within the
calibration uncertainty described in Section 6.5.5.
6.5 Report
6.5.1 Suppliers of CD-SEM magnification references
must report the following to the purchaser.
6.5.2 Wafer material for wafer-type magnification
references, or chip and stub materials for chip-type
magnification references: e.g., silicon chip, aluminum
stub, etc.
6.5.3 Type of reference patterns: e.g., dense lines, etc.
6.5.4 Figures and/or pictures of the reference patterns
with pattern dimensions which show top and/or cross
sectional view.
6.5.5 Calibration uncertainty of reference patterns:
e.g., line pitch 180
± 1 nm at 95% confidence level, etc.
6.5.6 Traceability and its certification organization,
and measurement method used for deciding the
calibration uncertainty of reference patterns: e.g., the
uncertainty was decided in terms of mean value and
variation obtained by means of a precise diffraction
angle measurement using a He-Cd laser beam of a 2
mm spot diameter, and was certified by NMI (National
Metrology Institute), etc.
NOTE 1: Traceability is the property of the calibration of the
reference patterns whereby it can be related to stated
references, usually national or international standards, through
an unbroken chain of comparison of all having stated
uncertainties. Refer to the international vocabulary of basic
and general items in metrology, ISO, 1993, 60 p., ISEN 92-67
01075-1.
6.5.7 Applicable magnification range: e.g., 50k×
200k×, etc.
6.5.8 Notices of caution for cleaning and maintenance:
e.g., whether cleaning is possible or not, and if possible,
the process of cleaning, procedures for storage, etc.