SEMI C10-0305 © SEMI 1998, 2005 1 SEMI C10-0305 GUIDE FOR DETERMINATION OF METHOD DETECTION LIMITS This guide was technicall y approved by the Global Liquid Chemicals Committee and is the direct responsibility of the Nor…

SEMI C2-95 © SEMI 1981, 1996 2
endpoint. Record the volume of base delivered (A).
Add 25 mL of Acetate Buffer for Hydrofluoric Acid
Determination and 1 g of Eriochrome Cyanine R
indicator mixture; mix thoroughly. With magnetic
stirring, titrate with the aluminum chloride solution to a
pink-purple endpoint. Record the volume of this
solution delivered (B).
soln. AlCl of B
2.001NaOH of A
Factor HF
3
××
=
N
4.4 Eriochrome Cyanine A Indicat or Mixture — Using
a mortar and pestle, thoroughly grind and mix 0.1 g of
Eriochrome Cyanine R (C.I. 43820) with 100 g of
potassium nitrate.
5 Safety
Because of the continuing evolution of safety
precautions, it is impossible in this publication to
provide definite statements relating to the safe handling
of individual etchants. The user is referred to product
labels, product and safety data sheets, government
regulations, and other relevant literature.
NOTICE: These standards do not purport to address
safety issues, if any, associated with their use. It is the
responsibility of the user of these standards to establish
appropriate safety and health practices and determine
the applicability of regulatory limitations prior to use.
SEMI makes no warranties or representations as to the
suitability of the standards set forth herein for any
particular application. The determination of the
suitability of the standard is solely the responsibility of
the user. Users are cautioned to refer to manufacturer’s
instructions, product labels, product data sheets, and
other relevant literature respecting any materials
mentioned herein. These standards are subject to
change without notice.
The user’s attention is called to the possibility that
compliance with this standard may require use of
copyrighted material or of an invention covered by
patent rights. By publication of this standard, SEMI
takes no position respecting the validity of any patent
rights or copyrights asserted in connection with any
item mentioned in this standard. Users of this standard
are expressly advised that determination of any such
patent rights or copyrights, and the risk of infringement
of such rights, are entirely their own responsibility.
Copyright by SEMI® (Semiconductor Equipment and Materials
International), 3081 Zanker Road, San Jose, CA 95134. Reproduction o
f
the contents in whole or in part is forbidden without express written
consent of SEMI.

SEMI C10-0305 © SEMI 1998, 2005 1
SEMI C10-0305
GUIDE FOR DETERMINATION OF METHOD DETECTION LIMITS
This guide was technically approved by the Global Liquid Chemicals Committee and is the direct
responsibility of the North American Liquid Chemicals Committee. Current edition approved by the North
American Regional Standards Committee on November 4, 2004. Initially available at www.semi.org January
2005; to be published March 2005. Originally published in 1998; previously published February 1999.
1 Purpose
1.1 To provide a minimal set of guidelines for the quantitative determination of a method detection limit (MDL)
from data supporting a SEMI Process Chemicals or Gases specification.
2 Scope
2.1 This guide applies to trace contaminants specified in SEMI Process Chemicals or Gases standards and
guidelines. All relevant trace contaminants should have an MDL determined from a regression analysis of a
calibration curve that is equal to, or less than, their specifications. This guide is intended for use in both establishing
new specifications within SEMI as well as verification of performance to SEMI specifications.
NOTICE: This standard does not purport to address safety issues, if any, associated with its use. It is the
responsibility of the users of this standard to establish appropriate safety and health practices and determine the
applicability of regulatory or other limitations prior to use.
3 Referenced Standards
3.1 SEMI Standard
SEMI C16 — Guide for Precision Reporting/Data Traceability
3.2 Other Document
1
Applied Regression Analysis, 3rd Edition, Norman R. Draper and Harry Smith, John Wiley and Sons, © 1998
NOTICE: Unless otherwise indicated, all documents cited shall be the latest published versions.
4 Terminology
None.
5 Boundary Conditions
NOTE 1: A series of boundary conditions is required of the data from which the MDL is to be determined from.
5.1 Data should be collected on at least two standards at different concentration levels, not including the blank.
5.2 The concentration levels investigated must either span the relevant specification or include it as a level. For this
purpose, the calibration range can be adjusted for instrument sensitivity by pre-concentration or dilution, if
applicable.
5.3 A minimum of three independent sample determinations should be taken on each concentration level of the
standard investigated.
5.4 Matrix Effects
5.4.1 The blank and the standards should be matrix-matched with the sample when testing a specification and run
under the same conditions, both during calibration and sample analysis.
5.4.2 If external standards in a different matrix are used for calibration, the user must compensate for matrix effects
using suitable internal standards.
1 John Wiley & Sons, Inc., 605 Third Avenue New York, NY 10158-0012, 212.850.6000

SEMI C10-0305 © SEMI 1998, 2005 2
5.5 Where applicable, this guide supersedes the earlier related SEMI Guide: SEMI C16.
5.6 The frequency of the MDL determination should be part of the quality assurance program for the laboratory.
6 Procedure
NOTE 2: The determination of the MDL should be performed under the same conditions that samples will normally be run. All
appropriate precautions to avoid contamination should be followed.
6.1 Reagents and Standards — Reagents and standards used should be of the same high-purity as used in the
analysis of samples. Accurate standard solutions of trace contaminants are needed for the preparation of the
calibration standards. Commercially available standard solutions are suitable. These solutions should be traceable to
NIST standards. Working standard solutions should be prepared on the day of use by dilution of the stock standard
solution. Standards and blanks should be matrix-matched to the sample to be analyzed.
6.2 Operating Conditions
6.2.1 Instrumental — Principles of instrumental analysis are provided within many monographs. No general
operating conditions are given here since these will vary with the instrument type and design; consequently, the
manufacturers manual should be followed.
6.2.2 Standard and Sample Preparation — With certain methods, the achievable MDL can be improved by pre-
concentration. Some methods require matrix dilution to run some analyses and may also affect the MDL.
6.2.3 Calibration — Calibrate the instrument for all specified trace contaminants using measurement and
preparation protocols identical to those used in sample preparation and measurement, wherever appropriate. The
calibration data structure used to determine the MDL should be as close to identical as possible to the data structure
normally used for calibration in the laboratory. Minimum data requirements should be determinations at, at least,
two concentration levels (standards), each independently sampled 3 times. The lowest concentration standard should
be at, or below, the specification and give a measurable signal. The blank may be used in the calculation of the MDL
if its variability is judged to be representative of that of a very low-level standard. All instrument response data used
in the analysis should be measurable. The calibration range can be adjusted for instrument sensitivity by pre-
concentration or dilution to match with how the samples are analyzed. Use the data collected to calculate the MDL
using the formulas in the next section.
6.3 Quantification of the MDL
6.3.1 The MDL is calculated by using the results from a regression analysis of the calibration data to obtain a 3
sigma equivalent (in probability) upper confidence limit of individual measurements. Two alternative computational
methods are provided for linear calibration models. One provides the MDL based on an ordinary least squares
(OLS) regression analysis; the other on a weighted least squares (WLS) based regression analysis. These two
computational alternatives do not provide the same MDL. The OLS algorithm is appropriate when signal
uncertainty remains relatively constant over the window of calibration; otherwise, the WLS algorithm is more
appropriate. A third alternative allows the use of similar regression-based approaches for MDL determination in the
presence of a more complex calibration model (e.g., nonlinear, multivariate), or when distributional assumptions of
the other alternatives are not appropriate. Many software packages perform regression analysis, and their use is
recommended.
6.3.2 Statistical Definitions
6.3.2.1 Method detection limit (MDL) — The level at which the errors in the measurement method become large
enough such that the preset maximum acceptable risk of seeing the quantified level, when none of the contaminant
in question is present in the sample, is exceeded.
6.3.2.2 risk — This guide uses 0.13% as the maximum acceptable risk. This is the probability of observing a given
result from a normal distribution that is 3 or more standard deviations above the average. Risk is set on a per-
contaminant basis.
6.3.2.3
calibration curve — An estimate of the relationship between concentration and instrument response (signal
intensity). A line, which has not been constrained to pass through the origin, is implicitly assumed in the
subsequently provided formulae.
6.3.2.4 regression analysis — A curve-fitting technique from which the calibration curve can be estimated.