SEMI G38-0996 © SEMI 1987, 2004 1 SEMI G38-0996 (Reapproved 1104) TEST METHOD FOR STILL- AND FORCED-AIR JUNCTION-TO- AMBIENT THERMAL RESIST ANCE MEASUREMENTS OF INTEGRATED CIRCUIT PACKAGES This test method was technicall…

100%1 / 7923
SEMI G37-88 © SEMI 1986, 1988 4
Figure 4
Package Warpage Warp in mils/Length in Inches = Warp Factor
Figure 5
Shoulder Bend Location
Figure 6
Lead Co-planarity
NOTICE: These standards do not purport to address safety issues, if any, associated with their use. It is the
responsibility of the user of these standards to establish appropriate safety and health practices and determine the
applicability of regulatory limitations prior to use. SEMI makes no warranties or representations as to the suitability
of the standards set forth herein for any particular application. The determination of the suitability of the standard is
solely the responsibility of the user. Users are cautioned to refer to manufacturer’s instructions, product labels,
product data sheets, and other relevant literature respecting any materials mentioned herein. These standards are
subject to change without notice.
The user’s attention is called to the possibility that compliance with this standard may require use of copyrighted
material or of an invention covered by patent rights. By publication of this standard, SEMI takes no position
respecting the validity of any patent rights or copyrights asserted in connection with any item mentioned in this
standard. Users of this standard are expressly advised that determination of any such patent rights or copyrights and
the risk of infringement of such rights, are entirely their own responsibility.
Copyright by SEMI® (Semiconductor Equipment and Materials
International), 3081 Zanker Road, San Jose, CA 95134. Reproduction o
f
the contents in whole or in part is forbidden without express written
consent of SEMI.
SEMI G38-0996 © SEMI 1987, 2004 1
SEMI G38-0996 (Reapproved 1104)
TEST METHOD FOR STILL- AND FORCED-AIR JUNCTION-TO-
AMBIENT THERMAL RESISTANCE MEASUREMENTS OF
INTEGRATED CIRCUIT PACKAGES
This test method was technically reapproved by the Global Assembly and Packaging Committee and is the
direct responsibility of the Japanese Packaging Committee. Current edition approved by the Japanese
Regional Standards Committee on July 23, 2004. Initially available at www.semi.org September 2004; to be
published November. Originally published in 1987; previously published September 1996.
1 Purpose
1.1 The purpose of this test is to determine the thermal
resistance of integrated circuit packages using thermal
test chips.
2 Scope
2.1 This test method deals only with junction-to-
ambient measurements of thermal resistance and limits
itself to still- and forced-air convection testing
environments.
NOTICE: This standard does not purport to address
safety issues, if any, associated with its use. It is the
responsibility of the users of this standard to establish
appropriate safety and health practices and determine
the applicability of regulatory or other limitations prior
to use.
3 Referenced Standards
3.1 SEMI Standards
SEMI G32 — Guideline for Unencapsulated Thermal
Test Chip
SEMI G42 — Specification for Thermal Test Board
Standardization for Measuring Junction-to-Ambient
Thermal Resistance of Semiconductor Packages
NOTICE: Unless otherwise indicated, all documents
cited shall be the latest published versions.
4 Terminology
4.1 Definitions
4.1.1 The following definitions and symbols shall
apply for the purpose of this test:
4.1.2 ambient temperature (T
A
, in degrees Celsius) —
the ambient temperature is the temperature of the air at
a specified location in the vicinity of the
microelectronic device under test (DUT).
4.1.3 junction temperature (T
J
, in degrees Celsius) —
the term is used to denote the temperature of the
semiconductor junction in the microcircuit in which the
major part of the heat is generated. For purposes of this
test, the measured junction temperature is only
indicative of the temperature in the immediate vicinity
of the element used to sense the temperature.
4.1.4 power dissipation (P
H
, in watts) — the heating
power applied to the device causing a junction-to-
reference point temperature difference.
4.1.5 temperature-sensitive parameter (TSP) — the
temperature-dependent electrical characteristic of the
junction under test which can be calibrated with respect
to temperature and subsequently used to detect the
junction temperature of interest.
4.1.6 thermal resistance — junction to specified
reference point, R
θJR
degrees Celsius/watt. The thermal
resistance of the microcircuit is the temperature
difference from the junction to some reference point in
the ambient divided by the power dissipation P
H
.
4.1.7 velocity (v
A
, in linear feet per minute (LFPM)) —
the velocity of the air at a specified location upstream
of the DUT.
5 Apparatus
5.1 The apparatus required for these tests shall include
the following as applicable to the specified test
procedures.
5.1.1 Thermocouple Material — Shall be copper
constantan (type T) or equivalent, for the temperature
range −100 to + 300°C. The wire size shall be no larger
than AWG size 30. The junction of the thermocouple
shall be welded to form a bead rather than soldered or
twisted. The accuracy of the thermocouple and
associated measuring system shall be ± 0.5°C.
5.1.2 Suitable Electrical Equipment — As required to
provide controlled levels of conditioning power and to
make the specified measurements. The instrument used
to electrically measure the temperature-sensitive
parameter shall be capable of resolving a voltage
change of 0.5 mV.
5.1.3 Controlled Temperature Chamber, Fluid Bath, or
Wind Tunnel — Capable of maintaining the specific
reference point temperature to within ± 0.5°C of the
reset (measured) value. Typical still-air enclosure and
SEMI G38-0996 © SEMI 1987, 2004 2
wind tunnel assemblies are presented for reference
purposes only.
5.1.4 Still-Air Enclosure Assembly — The microcircuit
shall be mounted in a cubic enclosure of not less than
0.028 m
3
(1.0 ft
3
). There shall be no radiation sources
other than the microcircuit under test in the enclosure.
The interior enclosure wall shall have a high reflectance
finish (emissivity < 0.1). The ambient temperature
should be measured by means of a thermocouple
mounted at a distance of approximately 2.54 cm (1.0 in)
beneath the DUT and 1.27 cm (0.5 in) from the test
board or socket.
5.1.4.1 The microcircuit shall be mounted in such a
manner that conduction cooling through the leads or the
test socket or both shall be small compared to the other
cooling mechanisms. No. 36 AWG wire should be
connected to the device test socket. The air flow (by
natural convection) should be unrestricted above and
beneath the device. An alternative approach would be
to use a mounting arrangement that approximates an
application environment. Such a reference mounting
configuration can be found in Specification, Thermal
Test Board Standardization for Measuring Junction-to-
Ambient Thermal Resistance of Semiconductor
Packages. Device mounting and test board positioning
inside the measuring chamber are depicted in Figure 1.
5.1.5 Wind Tunnel Assembly — A typical wind tunnel
design is shown with its dimensions in Figure 2.
5.1.5.1 The fan or blower should be placed
downstream of the DUT as depicted in Figure 2. A
static pressure differential measurement across a
calibrated nozzle is used to calculate the wind tunnel air
velocity, while a thermocouple is used to measure the
ambient temperature upstream of the DUT. Both of
these devices should be placed at a specified location
upstream of the DUT, as shown in Figure 2. The
thermocouple is located in the center of the wind
tunnel, 5.08 cm (2.0 in) from the test section and 2.54
cm (1.0 in) above the center plane of the DUT.
5.1.5.2 The microcircuit shall be mounted in such a
manner that conduction cooling through the leads or the
test socket, or both, shall be small compared to the
other cooling mechanisms. To minimize conduction
through the leads, No. 36 AWG wire should be
connected to the device test socket. The DUT should be
aligned so that the air front is parallel to the longer edge
of the package (air front hitting the package side). An
alternative approach would be to use a mounting
arrangement that approximates an application
environment. Such a referee mounting configuration
can be found in SEMI G42. Device mounting and test
board orientation inside the wind tunnel are depicted in
Figure 3.
5.1.5.3 Flow straighteners should be placed upstream
of the DUT, as shown in Figure 2. The flow
straighteners should provide a flat velocity profile
across the test section of the wind tunnel. This will
ensure that the DUT is exposed to a uniform velocity
across its entire cross section. A typical velocity profile
for well-developed turbulent flow is depicted in Figure
3.
5.1.5.4 Calibrated hot wire anemometer or nozzle with
suitable pressure gauges (p
1
and p
2
) for measuring the
pressure difference across the nozzle is used to
calculate the wind tunnel air velocity. When using the
nozzle, the pressure differential across the calibrated
nozzle is measured using a liquid monometer, typically
in inches of water.
5.1.5.5 Hot wire anemometer capable of verifying the
air velocity profile with an accuracy of ± 5%. The
velocity sensor should disrupt the airflow as little as
possible.
6 Procedure
6.1 Measurement of Wind Tunnel Air Velocity, v
A
—
The air velocity measurement techniques are direct
method and indirect method.
6.1.1 Direct Measurement of Wind Tunnel Air Velocity
— The direct method is based on a measurement by a
hot wire anemometer in wind tunnel.
6.1.2 Indirect Measurement of Wind Tunnel Air
Velocity, vA — The air velocity measurement technique
is an indirect method based on a pressure differential
measurement across a calibrated nozzle. The actual
velocity of interest is derived from a calibration curve
relating the pressure differential of the nozzle to the
volume (or mass) flow rate of the air through the
nozzle. The required air velocity (in linear feet per
minute) is then calculated by dividing the volume flow
rate (in cubic feet per minute) by the cross sectional
area at the entrance of the DUT test section (in square
feet). The volume flow rate shall be determined to
within an accuracy of ± 10%.
6.1.2.1 Air Flow Profile — Hot wire anemometer
should be used to verify that the flow profile of the air
front, measured within 5.08 cm (2.0 in.) of the test
section, does not vary by more than 10% across the
center 90% of the test section. The DUT and mounting
board/socket should not be in the test section when the
flow profile is determined.
6.2 Direct Measurement of Reference Point
Temperature, T
S
— For the purpose of measuring the
still-air junction-to-ambient microelectronic device
thermal resistance in a chamber, the ambient
temperature (T
R
= T
A
) should be measured with a