semi合集-English.pdf - 第7404页

SEMI MF1527-1104 © SEMI 2003, 2004 10 Section 7.2 (co nversion from dopant density to resistivity) of SEMI MF723, by solving the app ropriate equation given in this section for resistivity iteratively for the net carrier…

100%1 / 7923
SEMI MF1527-1104 © SEMI 2003, 2004 9
8.7 Wafer Conductivity Type and Surface Orientation
— For control of four-point probes or for calibrating
eddy current probes, conductivity type and surface
orientation do not influence the result. In this case
these parameters should be chosen so as to provide the
most uniform specimens available. For low resistivity
reference wafers, p-type (100) or (111) wafers are most
appropriate, while for high resistivity reference wafers,
n-type (neutron transmutation doped) wafers of either
orientation are preferred. For use in calibrating
mercury probes, the conductivity type of the reference
wafers must match that of the specimens to be tested.
For use in calibrating spreading resistance instruments,
the type and surface orientation of the reference wafers
must match those of the specimens to be tested.
NOTE 6: When uniformity considerations are the dominant
issue, the following should be kept in mind. At low
resistivity, wafers cut from Cz crystal are not only more
readily available but probably have better uniformity than
wafers cut from FZ crystals. Wafers cut from MCz crystals,
if available, may be an acceptable alternative to NTD wafers
for the higher resistivity ranges. In general, (100) crystals
have better radial resistivity uniformity than (111) crystals.
Also p-type crystals have generally better radial resistivity
uniformity than n-type crystals.
9 Preparation and Calibration of Resistivity
Reference Wafers
9.1 Wafer Preparation — Wafers should be sliced
from a qualified crystal section of the desired
resistivity, diameter, conductivity type, and orientation.
The thickness of the as-cut wafers should be at least
100 m larger than the desired finished thickness to
allow for removal of 25 m from each surface during
both etching and lapping. Etching immediately after
slicing in either KOH etching solution (see Section 5.4)
or etching solution (15 + 1) (see Section 5.5) is
recommended to remove damage introduced during
slicing. Subsequent lapping with 5 to 9-m aluminum
oxide abrasive is intended to provide both control of
thickness variation and a uniform matte surface.
9.1.1 The total thickness variation (TTV) of resistivity
reference wafers should be as small as possible. As a
minimum, the reference wafers should meet the TTV
requirement of ±1 % of the center-point thickness as
specified in SEMI MF84. For 625-m thick wafers,
this would require TTV 12.5 m. In most cases, it
should be possible to obtain lapped wafers with TTV
6 m over the entire wafer and TTV 3 m (± 0.25%
of the center-point thickness) over the region of the
wafer where the resistivity uniformity is determined
(see Section 8.5.1.2). Etched or single-side polished
wafers may have somewhat larger TTV.
9.1.2 The preferred surface finish is that obtained by
lapping with 5-m alumina powder as specified in
SEMI MF84; generally similar results can be obtained
by lapping with alumina powders up to 9 m. This
surface finish was originally chosen to provide good
probe-tip and specimen-surface wear together with
sufficient smoothness to reduce probe wander.
9
However, resistivity reference wafers intended for
calibration of mercury probe and spreading resistance
instruments must be polished. On the other hand, if
polished wafers are used for four-point probe
measurements, difficulties may be encountered with
accumulated probe damage and the resultant contact
quality degradation or surface charge, or both.
However, use of instruments calibrated with lapped
resistivity reference wafers for measuring polished
wafers could result in errors because the electrical
thickness may be different for lapped and polished
wafers of the same mechanical thickness.
NOTE 7: An unpublished roughness-thickness correlation
study has shown that the electrical thickness of lapped wafers
with root-mean-square (rms) surface roughness < 0.58 m on
both sides does not differ from the mechanical thickness by
more than 2.5 m.
9.2 Calibration
9.2.1 Determine that the candidate resistivity reference
wafer meets the resistivity uniformity requirements
appropriate to the intended application or applications
(see Sections 8.1, 8.4, and 8.5).
9.2.2 Determine the resistivity of the candidate
reference wafer at its center in accordance with SEMI
MF84 except that (1) thickness should be measured in
accordance with Section 8.6.2, (2) each determination
should consist of m measurements, where 6 m 10,
taken with the wafer rotated about (360/m) deg between
each measurement, and (3) wafers with a surface finish
that results in stable, reproducible measurements may
be used. If the measurement is made on an etched or
polished surface, equivalence with the result that would
have been obtained using a lapped surface must be
demonstrated.
9.2.3 Calculate the average resistivity corrected to
23°C in accordance with SEMI MF84 and the sample
standard deviation of the six to ten individual resistivity
measurements (corrected to 23°C).
9.2.4 If the resistivity reference wafer is to be used for
calibrating a mercury probe system, convert the
temperature-corrected average resistivity value to net
carrier density using the computational method given in
9 Hargreaves, J. K., and Millard, D., “The Accuracy of Four-probe
Resistivity Measurements on Silicon,” Brit. J. Appl. Phys. 13, 231–
234 (1962).
SEMI MF1527-1104 © SEMI 2003, 2004 10
Section 7.2 (conversion from dopant density to
resistivity) of SEMI MF723, by solving the appropriate
equation given in this section for resistivity iteratively
for the net carrier density.
9.2.5 Provide the following information with the
reference wafer:
9.2.5.1 Wafer identification (including source crystal
and position therein, orientation, conductivity type, and
dopant impurity),
9.2.5.2 Date of calibration,
9.2.5.3 Calibrating laboratory and operator,
9.2.5.4 Identification of instrumentation used,
9.2.5.5 Nominal wafer diameter, in mm,
9.2.5.6 Average wafer thickness, in mm,
9.2.5.7 Total thickness variation over the 38-mm
diameter central circle, in m,
9.2.5.8 Method for determining thickness and
thickness variation,
9.2.5.9 Nominal measuring current, in mA,
9.2.5.10 Average center-point resistivity, corrected to
23°C, in ·cm,
9.2.5.11 Number of resistivity measurements (m) per
determination, and
9.2.5.12 Standard deviation of corrected center-point
resistivity values, in ·cm.
9.2.6 If appropriate for the application also provide any
or all of the following:
9.2.6.1 Net carrier density, in cm
3
,
9.2.6.2 Diameter of the area over which thickness and
radial resistivity uniformity were obtained, if different
from 38 mm,
9.2.6.3 Measures of macro-scale and micro-scale radial
sheet resistance (resistivity) uniformity, including the
method or methods by which the uniformity was
determined,
9.2.6.4 Measure of axial resistivity uniformity
estimate, including a description of the procedure by
which the estimate was obtained, and
9.2.6.5 Average voltage-current ratio, corrected to
23°C for a four-point probe with ideal (equal) 1.59-mm
probe-tip spacing.
10 Application of Resistivity Reference
Materials
10.1 Calibration procedures for various resistivity
measurement equipment are given in the applicable
resistivity test method as summarized in Table 2.
10.2 For control of instruments for routine resistivity
measurements, a control chart for individuals with
moving range should be employed. In general, single
measurement determinations of resistivity are made for
routine purposes; therefore, the control should be
carried out on this basis. Established procedures should
be used for generating and maintaining the individuals
and moving range control charts, for determining the
existence of out-of-control conditions, and taking the
appropriate corrective action. A suggested procedure is
given in Section R1-4 for use in organizations without
previously established procedures.
11 Operator Training Requirements
11.1 Personnel responsible for selection and calibration
of resistivity reference wafers should be skilled in the
following areas:
11.1.1 Operation of four-point probe resistivity
measurement apparatus, including familiarity with
SEMI MF84 or DIN 50431,
11.1.2 Sources of radial and longitudinal resistivity
variations in silicon crystal,
11.1.3 Operation of resistivity mapping apparatus,
manual or automatic, as appropriate,
Table 2 Summary of Test Methods Applicable to Resistivity and Related Measurements
Measurement Equipment Applicable Test Method(s)
In-line Four-point Probe SEMI MF84 or DIN 50431
Non-contact Eddy-current Gage SEMI MF673, DIN 50445 or DIN 50447
Spreading Resistance Probe SEMI MF525 or SEMI MF672
Mercury Probe SEMI MF1392, SEMI MF1393 or DIN 50439
Dual-configuration Four-point Probe SEMI MF1529
SEMI MF1527-1104 © SEMI 2003, 2004 11
11.1.4 Operation of eddy current gage, spreading
resistance probe, and mercury probe, as required,
including familiarity with appropriate standard test
methods (see Section 10.1),
11.1.5 Generation and maintenance of
X
and s control
charts for measurement equipment together with
recognition of out-of-control conditions and the
appropriate corrective action procedures,
11.1.6 Measurement of wafer thickness and TTV, and
11.1.7 Care and use of resistivity CRMs.
11.2 Individuals responsible for preparation of
resistivity reference wafers should have a good
understanding of wafer shaping processes, especially
slicing and lapping.
11.3 Operators and other personnel associated with the
use of resistivity reference wafers for calibration or
control of resistivity measurement equipment used in
production or for other routine measurements, should
be well trained in the use and maintenance of the
equipment as well as in meticulous record keeping;
accurate records are essential to the integrity of the
calibration and control procedures. Familiarity with
statistical process control procedures including
generation and maintenance of moving range control
charts, recognition of out-of-control conditions, and the
appropriate corrective action chain is also required.
12 Keywords
12.1 certified reference materials; control chart; eddy
current gage; four-point probe method; mercury probe;
reference materials; resistivity; resistivity reference
wafer; semiconductor; sheet resistance; silicon wafers;
SPC; spreading resistance probe; Standard Reference
Materials