semi合集-English.pdf - 第3818页
injection line should be capped, excep t when performing an injection test. 7.3.5 Perform several tests with differing d irections of release relative to any opening or penetration in the enclosure. The location an d dir…

6 Summary of Method
6.1 A test is performed by releasing tracer gas at a
constant flow rate within an enclosure to simulate a
worst case leak and then measuring on the periphery of
the enclosure for the presence or absence of tracer. The
lack of measurable tracer indicates that the release of
potentially hazardous gases or vapors within the
enclosure at the tracer injection point(s) will not result
in their migration to the outside of the enclosure. Gas
samples are taken by means of disposable syringes,
sample bags, or sample vials. Gas samples are
analyzed by means of electron capture gas
chromatography.
7 Procedure
7.1 Test Design
7.1.1 Determine the type of enclosure to be tested, such
as: non-access, access, vacuum pump, equipment
cabinet, or other.
7.1.2 Determine the volume of the enclosure.
7.1.3 Measure the exhaust flow rate from the
enclosure.
7.1.4 Calculate the air changes per minute of the
enclosure by dividing the exhaust flow rate (Section
7.1.3) by the enclosure volume (Section 7.1.2).
7.1.5 Calculate the time at which the tracer
concentration in the enclosure will achieve approximate
equilibrium. Concentration equilibrium occurs when
the tracer concentration in the enclosure stops changing
as a function of time for a constant tracer release rate.
Divide 3 by the air changes per minute to establish this
time. Use this as the time to take the first sample after
initiating a test. Appendix 1 provides a derivation of
the equilibrium time.
NOTE 1: This test method is intended to test the containment
ability of the local exhaust system within an enclosure when
operated according to the manufacturer’s specifications. Thus,
testing should be performed with the local exhaust operating
under its manufacturer's recommended conditions.
7.1.6 If an excess flow control system is used,
determine the trip point for each hazardous gas used. If
different gases are used in the enclosure, the largest trip
point should be used to calculate the tracer release flow
rate.
7.1.7 If no excess flow system or flow-restricting
orifice is used, the maximum accidental release rate
must be calculated from the known maximum system
pressure and either valve coefficients or tubing ID. An
equation for flow through straight tubing is provided in
Appendix 2.
7.1.8 In the absence of guidance in the above two
sections, a standard test is the release of tracer gas
through 6.35 mm (0.25 inch) OD by 0.89 mm (0.035
inch) wall tubing at a rate of 28 standard liters per
minute (slpm) (1 scfm). Appendix 3 derives a general
equation, A3-7, that can be used when this is not the
case.
7.2 Reagents and Materials Use SF
6
diluted in an
inert gas, such as nitrogen or argon, as the tracer source,
to minimize measurement difficulties associated with
small leaks of pure SF
6
from the supply cylinder and its
associated piping.
7.3 Sampling
7.3.1 In selecting the location of samples collected
outside the enclosure, consider 1) potential leak points,
2) the direction of the release, and 3) laminar flow
characteristics in the area surrounding the enclosure.
Samples should be collected from all sides of the
enclosure, downstream in the prevailing room air flow,
and in the operating personnel occupancy areas.
7.3.2 The time required for the enclosure to reach
equilibrium should be considered when establishing the
time to begin sampling. The first sample after initiating
tracer flow should be taken at the enclosure equilibrium
time. Collect additional samples at 1 to 2 minute
intervals until the tracer source is shut off. One sample
should be taken 1 minute after the tracer source is shut
off. The test duration can be changed to accommodate
a particular test.
7.3.3 Collect background (baseline) samples from the
area surrounding the enclosure at predetermined
locations. When logistics permit, analyze the
background samples before releasing the tracer gas. If
background levels above approximately 1 ppb are
detected, evaluate the integrity of the SF
6
tracer
delivery system, and postpone the test until the
concentration is less than 1 ppb. Other sources of SF
6
in the immediate test area may also cause this
background.
NOTE 2: If testing is performed with an SF
6
background, the
background concentration must be measured and subtracted
from any subsequently measured tracer concentration value.
7.3.4 Release tracer gas within the enclosure being
tested by means of an injection manifold, shown
schematically in Figure 1. The tracer injection
manifold must be capable of measuring flow rates to an
accuracy of ± 5%. The tracer gas delivery line must be
routed into the enclosure and attached to a potential
leak point without violating the integrity of the
enclosure.
NOTE 3: To minimize tracer gas contamination of the area
surrounding an enclosure during a test, the end of the tracer
SEMI F15-93 © SEMI 1993, 2004 2

injection line should be capped, except when performing an
injection test.
7.3.5 Perform several tests with differing directions of
release relative to any opening or penetration in the
enclosure. The location and direction of the release
shall effectively simulate an actual gas release or
tubing/fitting failure within the enclosure. A worst case
failure can be simulated by locating the tracer injection
point at the potential leak location closest to a
penetration or opening within the enclosure with the
direction of tracer injection pointed directly at the
opening or penetration.
7.3.6 After initiation of tracer injection, collect grab air
samples from the area surrounding the enclosure at
predetermined times and locations. These samples
should be analyzed immediately after collection. If this
is not possible, they should be sealed. Label the
samples as to time and location. Samples may be taken
with: a) Containers that are non-absorbent, inert, and
that have low permeability (such as polyvinyl fluoride
film or polyester film sample bags or polyethylene,
polypropylene, nylon, or glass bottles) or b) disposable
syringes. Disposable syringes can be used to inject
samples into the gas chromatograph directly.
7.3.7 Record: a) the actual (measured) tracer gas
release rate, b) the actual SF
6
concentration in the tracer
gas being used, and c) the actual release time during a
test.
7.3.8 Collect air samples as described in Section 7.3.6,
and analyze them for the presence or absence of tracer
gas using a gas chromatograph. The measurement of a
non-zero tracer concentration in the area surrounding an
enclosure indicates incomplete containment of
contaminants within the enclosure.
7.3.9 Analyze samples according to ASTM E 260 and
ASTM E 697 in conjunction with the chromatograph
manufacturer's operating procedures. Samples may be
analyzed immediately after a test, or they may be stored
for future analysis. Experience has shown no
degradation of concentration in polypropylene syringes
when stored for several months as long as the needle or
syringe is plugged. Polypropylene syringes should be
discarded after one use to eliminate the possibility of
cross contamination of samples.
8 Calculations and Interpretation of Results
8.1 The maximum concentration of SF
6
measured in a
sample collected outside the enclosure is used to
calculate the Equivalent Release Concentration (ERC)
by the following formula:
ERC =
Conc. o
f
process gas
Measure
d
SF
6
Conc.
Tracer Injection gas conc.
8.1.1 The above formula assumes the process gas
tubing that fails is the same diameter as the tracer gas
injection tubing. Appendix 3 contains a general
equation that can be used when this is not the case.
8.2 Compare the Equivalent Release Concentration to
the hazardous process gas TLV or PEL, whichever is
lower. If the ERC is above the prescribed limits, the
enclosure is not considered to be acceptable; if the ERC
is less than or equal to the prescribed limit, the
enclosure is considered acceptable. SEMI S2
recommends appropriate control limits for an enclosure.
9 Reporting Results
9.1 Present all results in tabular form in a manner
which unambiguously notes those enclosure(s) that do
not satisfy the criteria of SEMI S2. The exhaust
operating conditions during each test must be provided
along with each measured ERC value.
9.2 Describe tracer gas injection points within
individual enclosures to detail location and proximity to
openings, penetrations, exhaust grillwork, access
panels, and other potential leakage sites, such that worst
case leak conditions have been simulated.
9.3 Describe the unit being tested by manufacturer,
model number, and serial number to provide
identification of the unit being tested.
9.4 Tabulate and record the name, supply
concentration, maximum flow rate, and TLV or PEL for
each hazardous gas or vapor used within each enclosure
being tested.
Figure 2
Schematic Drawing of Injection Manifold
SEMI F15-93 © SEMI 1993, 2004 3

10 Related Documents
29 CFR 1910.1000 Code of Federal Regulations,
Title 29, Superintendent of Documents, U.S.
Government Printing Office, Washington, D.C. 20402.
ACGIH American Conference of Governmental
Industrial Hygienists, 6500 Glenway, Building D-7,
Cincinnati, OH 45211-4438, (513)661-7881. Industrial
Ventilation - A Manual for Recommended Practice,
20th ed. ISBN: 0-936712-65-1.
American Industrial Hygiene Association 345 White
Pond Drive, Akron, OH 44320, (216)873-2442.
Workplace Environmental Exposure Level Guides.
ASTM American Society for Testing and Materials,
1916 Race Street, Philadelphia, PA 19103. ASTM
Standard E741-83 (Determining Air Leakage Rate by
Tracer Dilution).
BOCA Building Officials and Code Administrators
International Inc., 4051 West Flossmoor Road, Country
Club Hills, IL 60478-5795, (708)799-2300. National
Fire Prevention Code.
Grot, R.A. and P.L. Lagus “Applications of Tracer
Gas Analysis to Industrial Hygiene Investigations,”
Industrial Hygiene News, May 1991.
Orcutt, J.R. “Characterization of Hazardous Gas
Releases by Tracer Gas Simulation,” Hazardous
Assessment and Control Technology in Semiconductor
Manufacturing, ACGIH, Cincinnati, OH, 1988.
SBCCI Southern Building Code Congress
International, 900 Montclair Road, Birmingham, AL
35213-1206, (205)591-1853. Standard Fire Prevention
Code.
Tubby, R.L. “Tracer Gas Testing of Secondary
Exhaust Systems on Hazardous Gas Enclosures,” SSA
Journal, Vol. 5, June 1991.
UFC Uniform Fire Code. International Conference
of Building Officials and Western Fire Chiefs
Association, 5360 South Workman Mill Road, Whittier,
CA 90601. 1988 Edition, ISSN 0896-9736.
SEMI F15-93 © SEMI 1993, 2004 4