semi合集-English.pdf - 第4767页

SEMI C15-95 © SEMI 1995, 2002 4 of the moisture generator, and to its op erating temperature. If the background moisture concentration is B (ppb), the concentration d elivered by the moisture standard is C (ppb) and the …

100%1 / 7923
SEMI C15-95 © SEMI 1995, 2002 3
order to improve agreement with the primary
hygrometer.
6.2.3 Frequency of Hygrometer Validation — The
frequency of validation required for a secondary
hygrometer will vary dependent upon the performance
established for hygrometers of the same general type.
If three successive monthly validations show continuing
agreement (within 10%) with the primary hygrometer
without calibration adjustment, the validation interval
may be extended to three months. Similar extensions of
validation interval to six months and one year may be
made if no drift is apparent and no calibration
adjustments are necessary to bring the secondary within
10% of the primary at all points.
6.3 Validation Procedure for Moisture Standards
6.3.1 Validation Within the Range of an Available
Primary Hygrometer — If a primary hygrometer is
available, and its detection limit is no higher than the
lowest concentration at which it is desired to use the
moisture standard, this is the simplest case. Use the
standard to generate four concentration levels or two
per decade (whichever is less) in which the standard
will be used, half of the concentrations generated while
increasing concentration, half while decreasing,
including at least one comparison at the highest point of
the concentration range. The first and last comparisons
should be at the lowest point of the concentration range.
If the generated levels, calculated according to the basic
properties of the generator (e.g., weight loss of a
permeation tube) agree with the measured levels to
within 10%, the standard may be considered validated
with an accuracy of 10%. Higher validation accuracy
requires better agreement. It has been assumed that the
precision of the primary hygrometer, as determined by
a national standards laboratory, is much better than
10% and may be neglected for validation at the 10%
level. Validation at higher accuracy requires explicit
consideration of the certification precision.
If a standard only generates one level, connect it to the
hygrometer and allow to come to equilibrium.
Assuming the equilibrium level agrees with the
calculated level of the generator, disconnect the
hygrometer and bring it to equilibrium with either dry
nitrogen or ambient air. Then reconnect the moisture
standard and reestablish equilibrium between the
standard and the hygrometer. Disconnect again and
repeat a third time. If the three measurements agree
within 10% among themselves and with the generated
level, the standard is validated.
It is implicit in the above that the background moisture
level of the system be less than 50 ppb so that it can be
neglected for all measurements above 1 ppm with an
error no larger than 5%. We can usually neglect
background moisture in validating moisture standards
in this concentration range.
6.3.2 Validation Using Secondary Hygrometers — If
no primary hygrometer is available, then validation can
be performed using a secondary hygrometer. If the
lowest concentration to be generated using the standard
is no higher than the detection limit of the primary
hygrometer used for validation of the secondary
hygrometer, then the procedure is as above, except that
the secondary must first have been validated over the
entire concentration range which the standard will be
used to generate. (A secondary hygrometer and a
moisture standard can be validated simultaneously if
the primary and secondary hygrometers are connected
in parallel to the moisture standard.)
When assessing the accuracy of validation using a
secondary hygrometer, it is necessary to take into
account the precision of the comparison of the primary
hygrometer with the secondary hygrometer as well as
the precision of the comparison of the generator with
the secondary hygrometer.
7 Practice for Generating ppb Moisture
Standards
The detection limit of the primary hygrometer will
generally be on the order of 1 ppm, whereas it will
frequently be necessary to use the moisture standard at
concentrations below this level, by a factor of as much
as 10,000. Therefore, valid methods for reducing
concentration are necessary.
7.1 Dilution of Validated Moisture Standards — If a
moisture standard validated in accordance with Section
6.2 or Section 6.2.1 is diluted with a zero gas stream
satisfying the criteria of Section 5.3, then the resultant
diluted standard may be considered valid and may be
used for calibration of analyzers, such as APIMS.
7.2 Consideration of Background Moisture — In any
ppb or lower moisture measurement, it is necessary to
consider the background moisture level due to
outgassing in the sampling system and residual
moisture in the generator + analyzer. Often the
background will be sufficiently large that it must be
taken into account explicitly. The background moisture
level is the moisture concentration measured in the
absence of a moisture source (i.e., when the driest
carrier gas available is fed to the analyzer using the
same flow rate and flow path as is used for the sample
gas during analysis).
It should be noted that the importance of the
background level increases for measurements at low
concentrations, and that the background level is
sensitive to the materials of construction, to the design
SEMI C15-95 © SEMI 1995, 2002 4
of the moisture generator, and to its operating
temperature.
If the background moisture concentration is B (ppb),
the concentration delivered by the moisture standard is
C (ppb) and the dilution factor x is given by
x =
flow of moisture standard
total flow
(1)
then c, the delivered moisture concentration is given (in
ppb) by
c = (1
x
)B +
xC
(2)
The accuracy of calibration will reflect the combined
accuracy of the standard and the dilution system,
together with the accuracy of the background
determination; all must be taken into account in any
statement of analytical uncertainty.
7.3 Changing the Range of Performance of a Moisture
StandardIf a moisture standard is based on some
device which is believed to deliver moisture at a
constant rate, such as a permeation or effusion tube, the
concentration range delivered by the moisture standard
may be changed by delivering a substantially different
flow of carrier gas. This usually requires changing a
mass flow controller or other flow control device inside
the standard for a flow controller of a different range.
This may be done provided the new flow controller' s
performance is verified by comparison with a reliable
flow standard.
The definition of dilution factor is revised to
x =
flow generated by original flow controller
flow generated by new flow controller
(3)
The background, B, is determined by removing the
permeation or effusion tube from the system and
capping off its point of connection, or by preventing the
source from delivering moisture by some other means.
The same equation for c applies as does the statement
regarding analytical uncertainty.
7.4 Validation at Intermediate Concentrations — It
may be desirable to validate a moisture standard at a
concentration below the detection limit of the primary
hygrometer, if that standard is to be used for very low-
level generation. For example, if a standard based on a
permeation device is intended to generate levels
between 0.5 and 5 ppb at flows of 1 to 10 slm, it could
be validated in the 0.5 to 5 ppm range using a frost-
point hygrometer by replacing the flow controller with
one operating between 1 and 10 sccm
1
. However, such
a flow controller may not be readily available and is
difficult to calibrate in any case. Instead, it is preferable
to use a flow controller operating between 50 and 500
sccm to generate concentrations between 0.1 and 1.0
ppm together with a secondary hygrometer having a
detection limit no higher than 50 ppb. This procedure is
acceptable, provided that
the secondary hygrometer has been validated with
a primary hygrometer at least at one point in the
range to be used for validation of the moisture
standard,
the moisture concentrations generated by the
standard and that observed by the secondary
hygrometer agree throughout the range of
validation of the standard.
This approach could lead to errors if the hygrometer
and the standard each exhibited a deviation from
predicted performance, but these deviations would have
to be the same and cancel, which is extremely unlikely.
In the case of operation in this mode, it will be
necessary to know the background level of the system
and use this as an input to the calculation of generated
moisture levels. To continue the example of a
permeation based system, if the permeation rate
determined by weight loss is P (ng/min), the flow rate
of carrier gas is F (sccm), K is a factor to convert
ng/min to sccm moisture, and B is the experimentally
determined background, then the generated moisture
level, c
gen
, is calculated according to
c
gen
=
KP
F
+ B (4)
For other moisture generation systems, it is similarly
possible to calculate the generated moisture level while
taking into account the background level of the system.
This assumes that the background moisture
concentration is constant, which is only approximately
true. Thus, in applying this approach, in addition to the
sources of error mentioned above, the estimate of
analytical uncertainty must also include the variation in
background moisture.
8 Validation Precision and Accuracy
This procedure does not explicitly consider the
estimation of the accuracy of a moisture standard in
1 It is assumed that, although flow controllers are calibrated using a
standard, no unusual measures are used to enable the full accuracy of
calibration to be achieved over the entire range of the flow controller.
In practice this means that a flow controller has a usable dynamic
range of 10-100% of its flow rating.
SEMI C15-95 © SEMI 1995, 2002 5
detail, because such a calculation can be made by
applying procedures described elsewhere. The accuracy
required in validation will vary with the proposed
application of the moisture standard and may be left to
the discretion of the user of the procedure. It is
recommended that a propagation of errors calculation
be carried out to estimate the accuracy of the validation.
If no such calculation is made, and the minimum
criteria of the procedure are followed, it is estimated
that a validated moisture generator may output moisture
levels which differ from the true values by up to 20%.
NOTICE: SEMI makes no warranties or
representations as to the suitability of the standards set
forth herein for any particular application. The
determination of the suitability of the standard is solely
the responsibility of the user. Users are cautioned to
refer to manufacturer’s instructions, product labels,
product data sheets, and other relevant literature
respecting any materials mentioned herein. These
standards are subject to change without notice.
The user’s attention is called to the possibility that
compliance with this standard may require use of
copyrighted material or of an invention covered by
patent rights. By publication of this standard, SEMI
takes no position respecting the validity of any patent
rights or copyrights asserted in connection with any
item mentioned in this standard. Users of this standard
are expressly advised that determination of any such
patent rights or copyrights, and the risk of infringement
of such rights, are entirely their own responsibility.
Copyright by SEMI® (Semiconductor Equipment and Materials
International), 3081 Zanker Road, San Jose, CA 95134. Reproduction o
f
the contents in whole or in part is forbidden without express written
consent of SEMI.