semi合集-English.pdf - 第780页

SEMI E104-0303 © SEMI 2000, 2003 16 APPENDIX 3 AEROSOL GENERATION AND AEROSOL DILUTION NOTICE : The material in this appendix is an official part of SEMI E104 and was appro ved by full letter ballot procedures on July 28…

100%1 / 7923
SEMI E104-0303 © SEMI 2000, 2003 15
A2-4 Counting Efficiency
A2-4.1 The counting efficiency for a specific particle
size is defined as the ratio of the detected concentration
of particles to the concentration actually present. The
determination of the counting efficiency of an LPPD
requires particles of known size and concentration in
the aerosol. The counting efficiency is affected by
several factors.
First, the counting efficiency is dependent on a
specific particle size.
The effect of particle concentration is addressed
under zero counting and coincidence.
Due to inhomogeneous light intensity within the
detection area of the particle counter, not all small
particles near the lower detection limit of the
instrument are detected.
If the sampling flow is not completely contained
within the defined detection area, then some
portions of the aerosol will not be counted.
A2-4.2 Ideally, the counting efficiency of a particle
counter covering 100% of the line cross-section area as
detection area would consist of a step function at the
point of the LDL. Real particle detectors have a gradual
transition (efficiency curve) instead of a step function.
The point with 50% detection probability of all particles
of a given size moving through the detection area is
often used as a reference point. The corresponding
particle size is called the minimum detectable particle
size (see sensitivity). If the slope of the efficiency curve
for two detectors with the same specified 50%
efficiency point is different, the two detectors may
show different particle counts for the same polydisperse
aerosol. To define an efficiency curve, more than one
particle size, e.g. 5 particle sizes around the LDL
covering 0% to 100% efficiency, should be used.
Acceptable counting efficiency for single particle
counting instruments is 50% ± 20% at the minimum
detection size and 100% ± 10% for all particles larger
than 1.5 times the minimum size.
A2-4.3 The detection area of most LPPDs comprises
only a small portions of the cross-section area of the
line. Therefore, it is not possible to specify a 50%
efficiency point. For this, the commonly used definition
for sensitivity could not be used for such instruments.
A2-4.4 The actual particle concentration is determined
with a reference particle counter with a 100% counting
efficiency for the employed particle size. It is important
that there are no differences between the concentration
within the detection area of the LPPD and the
concentration monitored by the RPC.
A2-5 Sensitivity
A2-5.1 The particle size corresponding to the point
with 50% counting efficiency of the particles of a given
size is defined as the minimum detectable particle size.
This is valid only for particle counters capable to reach
100% counting efficiency for a specific particle size.
This does not apply to most LPPDs.
A2-6 Particle Concentration Effects
A2-6.1 All optical particle detectors are able to operate
accurately only within a limited range of particle
concentration. It is not the scope of this document to
define how this limit is measured or how to verify the
specification.
A2-6.2 High Concentration Effects
A2-6.2.1 The upper limitation of the particle
concentration is chiefly based on coincidence effects.
A2-6.2.2 One effect is optical coincidence. If the
particle concentration is too high, more than one
particle is present within the detection area of the
instrument. The reported particle concentration will be
less than the true value, and the reported particle size
distribution will be shifted towards the indication of
larger particles than in reality.
A2-6.2.3 The other effect is electronic coincidence. It
is defined as the inability of the electronic pulse
processing system to detect and size individual pulses
that are too closely spaced. If there are so many pulses
that they cannot completely return to the baseline, they
become superimposed. Electronic saturation occurs.
Electronic coincidence introduces errors in both particle
size and counts as well. Normally, this problem is not
critical for modern electronic systems. Optical
coincidence can become a problem at particle levels
well below the point where electronic saturation occurs.
A2-6.2.4 Another error can occur when high
concentrations of particles just smaller than the lower
detection limit of the instrument are present. Even
though no single one of these particles will be detected,
scattering-light levels from these particles can increase
the background optical noise level. So, errors might be
produced in particle count data in the lower particle size
ranges reported by the counter.
A2-6.3 Low Concentration Effects
A2-6.3.1 It is obvious that it is necessary to collect
sufficient data to determine number and size of particles
within an acceptable confidence limit.
SEMI E104-0303 © SEMI 2000, 2003 16
APPENDIX 3
AEROSOL GENERATION AND AEROSOL DILUTION
NOTICE: The material in this appendix is an official part of SEMI E104 and was approved by full letter ballot
procedures on July 28, 2000 by the European Regional Standards Committee.
A3-1 Monodisperse Particle Suspension
A3-1.1 Monodisperse suspensions of particles (PSL)
are available in sizes from 0.02 µm. These particles are
grown by emulsion polymerization and are stabilized in
aqueous suspensions with an anionic surfactant. They
carry a negative charge which contributes to their
stability. The suspension normally contains 10% solids
and 2% dissolved, highly viscous stabilizer. The solid
contains up to 8% emulsifier and inorganics.
A3-1.2 After evaporation, these solids and the
stabilizer will both increase the diameter of the single
particle and generate a residual nucleus out of an empty
droplet. These residual particles are called secondary
aerosol. The diameter of residual nucleus could be up to
0.25 µm according to the PSL size and dilution of the
suspension. Most of these secondary aerosol particles
might be too small to be detectable with commonly
used particle detectors, but they will have an influence
on the noise level because of there high number.
A3-1.3 Out of a high-concentration suspension it is not
possible to aerosolize only individual particles.
Droplets containing more than one of the suspended
particles will become undesirable agglomerates upon
evaporation.
A3-1.4 For this, an adequate dilution of the suspension
is necessary to avoid the formation of aggregates and
the enlargement of the particles.
A3-2 Generation and Dilution of PSL
Suspension
A3-2.1 A diluted PSL suspension for particle
generation in an aerosol generator can be created by the
following procedure:
Shaking and/or ultrasonic treatment of the PSL
bottle distributed by the PSL manufacturer.
Placing one drop of PSL in one liter of deionized
or destilled water or Isopropanol. The water could
be cleaned using a filter with a pore size no more
than 10% of the size of the particles being used.
Shaking and/or ultrasonic treatment of the
suspension to disperse the particles.
A3-2.2 The suspension will probably need to be diluted
further to provide a required particle concentration.
This is necessary to avoid the formation of
agglomerates and to restrict the size of the secondary
aerosol particles. An equation to calculate the required
particle concentration is given in “The Generation of
Aerosols of Fine Particles” by O. Raabe (see Section
11).
A3-2.3 The particle suspensions distributed by the PSL
manufacturer shows no detectable variation in particle
characteristics when they are stored in a cool place over
the years. Diluted suspensions for atomization feature
an aging process. Diluted suspensions of polymer
spheres smaller than 1 µm should not be stored for
more than one week.
NOTE A3-1: Care should be taken to avoid contamination of
the polymer spheres and the suspension.
A3-3 Aerosol Generation
A3-3.1 The diluted suspension is nebulized in the
aerosol generator. Great account is taken of
concentration and size distribution of the generated
droplets and of the volume flow rate V´
G
. To achieve a
constant aerosol production, the volume flow rate and
the droplet size distribution should be constant and
independent of the supplies of suspension in the
reservoir. The variation in particle concentration should
be no more than 10% as measured by the reference
particle counter over a time period of 15 minutes or
more.
NOTE A3-2: Care should be taken to avoid contamination of
the aerosol generator.
A3-4 Aerosol Drying
A3-4.1 The water of the generated droplets will
completely evaporate when the relative humidity of the
aerosol flow behind the nebulizer is lower than 70%.
The volume flow of the suspension in the nozzle
increases the humidity of the filtered compressed air
which has a relative humidity of 10%15% in spite of
predrying. The use of a diffusion dryer or another
appropriate instrument allows a higher relative
humidity of the aerosol flow behind the aerosol
generator. The diffusion dryer uses silica gel desiccant
to remove the moisture. The desiccant shall either be
new or freshly regenerated. It is not wise to get the
aerosol in direct contact with the desiccant. Polymer
spheres will be precipitated and additional particles out
of the desiccant will change the particle size
distribution of the aerosol.
SEMI E104-0303 © SEMI 2000, 2003 17
A3-5 Aerosol Neutralization
A3-5.1 When dispersing the suspension, the particles
are charged. This surface charge should be removed
after drying the aerosol flow to avoid electrostatic
interactions with each other or the line walls. With the
help of the discharging distance of an electrostatic
neutralizer, the aerosol is exposed to a bipolar ion
source. Some electrostatic neutralizers may produce a
large number of ultra-fine particles which will combine
with the calibration aerosol and affects the signal-to-
noise-ratio of the LPPD and the reference particle
counter.
A3-6 Aerosol Size Separation
A3-6.1 When calibrating with ultra-fine particles, a
particle size separation might be necessary to remove
the residue particles and agglomerates. An electrostatic
classifier is most effective for particle sizes less than 1
µm. This instrument electrically charges the incoming
aerosol. With electrostatic deflection, it separates
selected particles of one mobility. The residue particles
and the larger aggregate particles would be stripped out
of the particle stream consisting of the desired polymer
particles. To charge the aerosol, the electrostatic
classifier uses a radioactive neutralizer. The aerosol
exiting the instrument will contain singly charged
particles. This low charge level makes the use of an
additional aerosol neutralizer unnecessary.
Nevertheless, conductive tubing should be used
between the aerosol generator and the LPPD and
reference particle counter to minimize electrostatic
particle loss.
A3-7 Aerosol Dilution
A3-7.1 The particle concentration of the aerosol
generated by the nebulizer might be too high. The
aerosol should be further diluted to achieve the required
concentration and to avoid coincidence errors. The
dilution could be executed with a mixing chamber in
which the aerosol is mixed with zero gas. The internal
chamber pressure should be stable and very close to
ambient atmospheric pressure at operational flow rates.
The spatial particle distribution in the exiting aerosol
should be as homogenous as possible.