POTASSIUM HYDROXIDE, 45% SOLUTION SEMI C40-0699 © SEMI 19991
SEMI C40-0699
SPECIFICATION FOR POTASSIUM HYDROXIDE, 45% SOLUTION
This specification was technically approved by the Global Process Chemicals Committee and is the direct
responsibility of the North American Process Chemicals Committee. Current edition approved by the North
American Regional Standards Committee on April 23, 1999. Initially available on SEMI OnLine May 1999;
to be published June 1999.
1 Purpose
1.1 The purpose of this document is to standardize
requirements for 45% potassium hydroxide solution
used in the semiconductor industry and testing
procedures to support those standards. Test methods
have been shown to give statistically valid results. This
document also provides guidelines for grades of
potassium hydroxide, 45% solution for which a need
has been identified. In the case of the guidelines, the
test methods may not have been statistically validated
yet.
2 Scope
2.1 The scope of this document is all grades of 45%
potassium hydroxide solution used in the
semiconductor industry.
3 Limitations
3.1 None.
4 Referenced Documents
SEMI C1 Specifications for Reagents
5 Terminology
5.1 None.
6 Physical Property (for information only)
6.1 Not applicable.
7 Requirements
7.1 The requirements for 45% potassium hydroxide
solution for Grade 1 are listed in Table 1.
8 Grade 1 Procedures
NOTE 1: Each laboratory is responsible for verifying the
validity of the method within its own operation.
8.1 Assay
Accurately weigh 70 to 80 g of sample in
a closed weighing bottle. Transfer to a 500 mL
volumetric flask, add in carbon dioxide-free water, cool
to 25°C, dilute to the mark with carbon dioxide-free
water, and mix thoroughly. Transfer 25.0 mL of this
solution to a 500 mL glass-stoppered flask, dilute with
175 mL of carbon dioxide-free water, add 5 mL of
barium chloride reagent solution, shake and allow to
stand for five minutes. Add 0.15 mL of
phenolphthalein indicator solution and titrate with
standardized 1 N hydrochloric acid. (Save this solution
for the potassium carbonate test.)
(g) Sample ofWeight
112.2 HCl of N mL
%Assay
××
=
8.2 Potassium Carbonate (K
2
CO
3
) To the above
titrated solution, add 0. 15 mL of methyl orange
indicator and continue the titration with 1 N
hydrochloric acid, making sure that no insoluble
carbonate remains. The additional hydrochloric acid is
equivalent to that consumed by the potassium
carbonate.
(g) Sample ofWeight
138.2 HCl of N mL
COK % 32
××
=
8.3 Ammonium Hydroxide Precipitate Weigh 40 g
of sample into a 250 mL beaker and dissolve in about
100 mL of water.
8.3.1 Cautiously add 26 mL of sulfuric acid to 30 mL
of water in a 150 mL glass beaker, cool; then cautiously
add the mixture to the solution of the sample, and
evaporate to dense fumes. Cool, dissolve the residue in
130 mL of hot water, and add ammonium hydroxide
until the solution is just basic to methyl red.
8.3.2 Heat to boiling, filter, wash with hot water and
ignite. The weight of the residue should not exceed 2
mg.
8.4 Insoluble Matter Dilute 100 g of sample in 250
mL of carbon dioxide-free water in a 500 mL stoppered
flask and cool to 25°C. The solution should be clear
and colorless. Filter through a tared, medium porosity
filtering crucible and wash thoroughly with hot water.
Dry the crucible at 105°C for 1 hour, cool and weigh
the residue. The weight of the residue should not
exceed 5.0 mg.
8.5 Stock Solution for Sections 8.6 through 8.13 —
Add 100 g of sample to 250 mL of carbon dioxide-free
water and dilute with such water to 500 mL. One mL
contains 0.20 g of sample. Store in a plastic container.
8.6 Nitrogen Compounds (as N) Dilute 20 mL of
the stock solution with 50 mL of water in a flask

SEMI C40-0699 © SEMI 1999 POTASSIUM HYDROXIDE, 45% SOLUTION2
connected through a spray trap to a condenser, the end
of which dips beneath the surface of 10 mL of 0.1 N
hydrochloric acid. For the standard, take 50 mL of
water in a similar flask, and add 10 mL of the stock
solution and 1 mL of the nitrogen standard solution
(0.01mg N/mL). To each flask add 0.5 g of aluminum
wire in small pieces, allow to stand for 1 hour, and
slowly distill to about 35 mL. To each distillate add 2
mL of freshly boiled 10% sodium hydroxide reagent
solution, dilute with water to 50, add 2 mL of Nessler
reagent, and mix. Any color in the solution of the
sample should not exceed that in the standard.
8.7 Chloride Dilute l0 mL of the stock solution
with water to 50 mL. Filter, if necessary, through a
chloride-free filter. To 10 mL of this solution, add 5
mL of water, 2 mL of nitric acid, and 1 mL of silver
nitrate reagent solution. Any turbidity developed
should not exceed that produced by 0.02 mg of chloride
(CI) ion in an equal volume of solution containing the
quantities of reagent used in the test.
8.8 Phosphate To 50 mL of the stock solution, add
15 mL of hydrochloric acid and evaporate to dryness on
a steam bath. Dissolve the residue in 25 mL of 0.5 N
sulfuric acid. Add 1 mL of ammonium molybdate
reagent solution and l mL of p-(methylamino)phenol
sulfate reagent solution. Allow the solution to stand for
2 hours at room temperature. Any blue color developed
should not exceed that produced when 0.02 mg of
phosphate (PO
4
) ion is treated as the sample.
8.9 Sulfate To 20 mL of the stock solution, add 5
mL of hydrochloric acid and evaporate to dryness on a
steam bath. Dissolve the residue in 20 mL of water and
evaporate again to dryness. Dissolve the residue in 20
mL of water, filter if necessary, and add 1 mL of dilute
hydrochloric acid (1+19) and 1 mL of barium chloride
reagent solution. Allow to stand for 10 minutes. Any
turbidity developed should not exceed that produced
when 0.04 mg of sulfate (SO
4
) ion is treated as the
sample.
8.10 Heavy Metals (as Ag) To 60 mL of the stock
solution cautiously add 15 mL of nitric acid. For the
standard add 0.05 mg of silver ion (Ag) to 10 mL of the
stock solution and cautiously add 15 mL of nitric acid.
Evaporate both solutions to dryness over a low flame or
on an electric hot plate. Dissolve each residue in about
20 mL of water, filter if necessary through a chloride-
free filter, and dilute with water to 25 mL. Adjust the
pH of the standard and sample solutions to between 3
and 4 (using a pH meter) with 1 N acetic acid or
ammonium hydroxide (10% NH
3
). Dilute each solution
with water to 40 mL and stir. Add 10 mL of freshly
prepared hydrogen sulfide water to each solution and
stir again. Any color developed in the sample should
not exceed that in the standard.
8.11 Iron Neutralize l0 mL of the stock solution
with hydrochloric acid, using a phenolphthalein
indicator solution, add 2 mL of HCI in excess, and
dilute with water to 50 mL. Add 30 to 50 mg of
ammonium peroxydisulfate crystals and 3 mL of
ammonium thiocyanate reagent solution. Any red color
developed should not exceed that produced when 0.01
mg of iron (Fe) ion is treated as the sample.
8.12 Nickel Dilute 20 mL of the stock solution to
50 mL with water and neutralize with hydrochloric acid
using a pH meter. Dilute to 85 mL with water and add
ammonium hydroxide to make the solution barely basic
(pH 8). Add 5 mL of bromine water and 5 mL of
alcoholic 1% dimethylgy1oxime solution. Any red
color developed should not exceed that produced when
0.02 mg of nickel (Ni) ion is treated as the sample.
8.13 Sodium Dilute 10 mL of the stock solution
with water to 100 mL to form the sample solution. To
prepare the standard solution, add 0.6 mg of sodium
(Na) ion to 10 mL of the stock solution and dilute with
water to 100 mL.
8.13.1 Using flame emission spectroscopy, observe
the emission of the standard solution at the 589 nm
sodium line. Observe the emission of the sample
solution at the 589 nm sodium line and also at a
wavelength of 580 nm. The difference between the
intensities observed for the sample solution at 580 nm
and 589 nm should not exceed the difference observed
at 589 nm between the sample solution and standard
solution (see SEMI Cl, Section 3.7, Flame Emission
Spectroscopy).
9 Grade 2 Procedures
9.1 This section does not apply to this chemical.
10 Grade 3 Procedures
10.1 This section does not apply to this chemical.
11 Grade 4 Procedures
11.1 This section does not apply to this chemical.
12 Grade 5 Procedures
12.1 This section does not apply to this chemical.
13 VLSI Grade Procedures
13.1 This section does not apply to this chemical.
14 Tier A Procedures
14.1 This section does not apply to this chemical.

SEMI C40-0699 © SEMI 19993
15 Tier B Procedures
15.1 This section does not apply to this chemical.
16 Tier C Procedures
16.1 This section does not apply to this chemical.
17 Tier D Procedures
17.1 This section does not apply to this chemical.
Table 1 Impurity Limits and Other Requirements for Potassium Hydroxide, 45% Solution
Previous SEMI Reference # --
Grade 1
(Specification)
Assay (KOH) 45.0 - 47.0%
Insoluble Matter 50 ppm max
Nitrogen Compounds (as N) 5 ppm max
Ammonium Hydroxide Precipitate 50 ppm max
Chloride (Cl) 50 ppm max
Phosphate (PO
4
) 2 ppm max
Potassium Carbonate (K
2
CO
3
)0.5% max
Sulfate (SO
4
) 10 ppm max
Heavy Metals (as Ag) 5 ppm max
Iron (Fe) 5 ppm max
Nickel (Ni) 5 ppm max
Sodium (Na) 0.03% max
NOTICE: These standards do not purport to address
safety issues, if any, associated with their use. It is the
responsibility of the user of these standards to establish
appropriate safety and health practices and determine
the applicability of regulatory limitations prior to use.
SEMI makes no warranties or representations as to the
suitability of the standards set forth herein for any
particular application. The determination of the
suitability of the standard is solely the responsibility of
the user. Users are cautioned to refer to manufacturer’s
instructions, product labels, product data sheets, and
other relevant literature respecting any materials
mentioned herein. These standards are subject to
change without notice.
The user’s attention is called to the possibility that
compliance with this standard may require use of
copyrighted material or of an invention covered by
patent rights. By publication of this standard, SEMI
takes no position respecting the validity of any patent
rights or copyrights asserted in connection with any
item mentioned in this standard. Users of this standard
are expressly advised that determination of any such
patent rights or copyrights, and the risk of infringement
of such rights, are entirely their own responsibility.